CVE-2019-25723
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Weakness
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.